Influence factor
|
Test item
|
Base condition
|
Working condition
| |
Environment temperature
|
All parameters
|
(23±2)°C
|
-10°C~ 40°C
| |
Relative humidity
|
All parameters
|
40%~ 60%
|
<80%
| |
Phase-to-neutral voltage
|
All parameters
|
(100±1%)V
|
1.0V~ 1000V
| |
Phase-to-phase voltage
|
True RMS phase-to-phase voltage
|
(200±1%)V
|
1.0V~ 2000V
| |
Current
|
True RMS current
|
(5±1%)A
|
10mA~ 6000A
| |
Network frequency
|
All parameters
|
50Hz±0.1Hz
|
40Hz~ 70Hz
| |
Phase shift
|
Active power and active energy
|
Cosφ=1
|
Cosφ: 0.2~ 1.0
| |
Reactive power and reactive energy
|
Sinφ=1
|
Sinφ: 0.2~ 1.0
| ||
Harmonic
|
All parameters
|
<0.1%
|
0.0%~ 100%
| |
Voltage unbalance
|
All parameters
|
<10%
|
0.0%~ 100%
| |
Working voltage of device
|
All parameters
|
DC9.8V±0.1V
|
DC9.5V~ 10.5V
| |
External electric field,
magnetic field |
All parameters
|
Should be avoided
| ||
Measured lead position
|
Measured related parameters of current
|
Tested wire at the center of clamp.
|
Influence factor
|
Test item
|
Base condition
|
Working condition
| |
Environment temperature
|
All parameters
|
(23±2)°C
|
-10°C~ 40°C
| |
Relative humidity
|
All parameters
|
40%~ 60%
|
<80%
| |
Phase-to-neutral voltage
|
All parameters
|
(100±1%)V
|
1.0V~ 1000V
| |
Phase-to-phase voltage
|
True RMS phase-to-phase voltage
|
(200±1%)V
|
1.0V~ 2000V
| |
Current
|
True RMS current
|
(5±1%)A
|
10mA~ 6000A
| |
Network frequency
|
All parameters
|
50Hz±0.1Hz
|
40Hz~ 70Hz
| |
Phase shift
|
Active power and active energy
|
Cosφ=1
|
Cosφ: 0.2~ 1.0
| |
Reactive power and reactive energy
|
Sinφ=1
|
Sinφ: 0.2~ 1.0
| ||
Harmonic
|
All parameters
|
<0.1%
|
0.0%~ 100%
| |
Voltage unbalance
|
All parameters
|
<10%
|
0.0%~ 100%
| |
Working voltage of device
|
All parameters
|
DC9.8V±0.1V
|
DC9.5V~ 10.5V
| |
External electric field,
magnetic field |
All parameters
|
Should be avoided
| ||
Measured lead position
|
Measured related parameters of current
|
Tested wire at the center of clamp.
|
Power supply
|
Rechargeable lithium-ion battery packs 9.6V, backup charger.
|
Battery indicator
|
Battery symbol shows dump energy. When the voltage is too low, automatic shutdown after the 1 minute.
|
Power consumption
|
Current consumption of normal test 490 mA, continuous working for 10 hours.
|
Display mode
|
LCD color screen, 640dots×480dots, 5.6 inches, display domain: 116mm×88mm.
|
Size of clamp
|
FR008 small sharp current clamp: 8mm×15mm;
FR020 circle current clamp: 20mm×20mm; FR050 circle current clamp: 50mm×50mm. FR300R Flexible Coil Current Sensor (with Integrator) : Ф300mm |
Instrument dimensions
|
L×W×H: 277.2mm × 227.5mm × 153mm.
|
Number of channels
|
4U/4I.
|
Phase-to phase voltage
|
1.0V~2000V.
|
Phase-to-neutral voltage
|
1.0V~1000V.
|
Current
|
FR008 current clamp: 10mA~10.0A;
FR020 current clamp: 0.10A~100A; FR050 current clamp: 1.0A~1000A; FR300R Flexible Coil Current Sensor (with Integrator) : 10A ~ 6000A |
Frequency
|
40Hz~70Hz.
|
Parameters of electricity
|
W, VA, Var, PF, DPF, cosφ, tanφ.
|
Energy parameters
|
Wh, Varh, Vah.
|
Harmonic
|
Order 0~50.
|
Total harmonic distortion
|
Order 0~50, each phase.
|
Expert mode
|
Yes.
|
Number of Transient records
|
150 sets.
|
Voltage flicker
|
Yes.
|
Starting current mode
|
Yes, 100 seconds.
|
3 phases unbalance
|
Yes.
|
Record
|
300 days(simultaneous recording 20 parameters, record one point every 5 seconds).
|
Min/Max recorded value
|
Measurement of maximum and minimum values over a period of time.
|
Alarm
|
40 different types of parameter selections, 12,800 sets alarm logs.
|
Peak
|
Yes.
|
Phasor diagram display
|
Automatically.
|
Capacity of snapshots
|
60.
|
Menu language
|
English/Chinese.
|
Communication
|
USB.
|
Automatic switching off
|
When an alarm campaign is initiated or a search for transients, an inrush current capture, or a trend recording is pending or in
progress, the device is not automatic switching off. |
In other test mode, 15 minutes without keystrokes, automatic shutdown after prompt 1 minute.
| |
Backlight function
|
Yes, suitable for use at night and dark place.
|
Weight
|
Host: 1.6kg (with battery).
|
FR008 small sharp current clamp: 168g×4;
| |
FR020 circle current clamp: 252g×4;
| |
FR050 circle current clamp: 463g×4;
| |
FR300R Flexible Coil Current Sensor (with Integrator) : 280g×4;
| |
Test wires and power adapter: 800g;
| |
Total weight: about 10.8kg (with packaging).
| |
Length of voltage test wire
|
3m.
|
Length of current sensor wire
|
2m.
|
Working temperature and humidity
|
-10°C~40°C; below 80%Rh.
|
Storage temperature and humidity
|
-10°C~60°C; below 70%Rh.
|
Input impedance
|
Input impedance of test voltage: 1MΩ.
|
Withstand voltage
|
Withstand 3700V/50Hz sinusoidal AC voltage for 1 minute between instrument wiring and shell.
|
Insulation
|
Between instrument wiring and shell ≥10MΩ.
|
Structure
|
Double insulation, with insulation vibration-proof sheath.
|
Safely rules
|
IEC 61010 1000V Cat III / 600V CAT IV, IEC61010-031, IEC61326, Pollution degree: 2.
|